Infection Control · single question
Which statement about biological (spore) monitoring of sterilizers is NOT true?
Show answer and explanation
Correct answer: C. Spore testing should be performed at least once a month
- A. ❌ True. Geobacillus stearothermophilus is the indicator organism for steam and chemical vapor; Bacillus atrophaeus is used for dry heat and ethylene oxide.
- B. ❌ True. Every load containing an implantable device is spore-tested, and the device is quarantined until the result is known.
- C. ✅ CORRECT — This is the FALSE statement. The CDC recommends biological monitoring AT LEAST WEEKLY, not monthly. Monthly testing would leave up to four weeks of instruments unverified after a sterilizer failure.
- D. ❌ True. After a positive (failed) spore test, items processed since the last negative test are recalled and reprocessed.
Note: Monitoring layers: mechanical (gauges) + chemical (indicators, every package) + biological (spores, weekly). Only the biological test proves sterilization.
Practice Infection Control questions Infection Control study notes