Per CDC guidelines, autoclave biological spore testing should be performed at what minimum frequency?
Show answer and explanation
Correct answer: C. At least weekly (with every load containing implantable items)
- A. Testing with each individual cycle is not required for non-implantable items.
- B. Monthly testing alone is insufficient per CDC standards — weekly is the minimum.
- C. Correct. The CDC recommends at least WEEKLY biological monitoring of heat sterilizers (autoclaves) using Geobacillus stearothermophilus spore strips/vials. Additionally, biological monitoring should be performed with EVERY load containing implantable items (items placed in tissue/bone). Defective pouches or failed spore tests require removing the sterilizer from service.
- D. Daily testing is not required by current CDC guidelines (it would be impractical and cost-prohibitive).