Which of the following statements about biological spore testing (autoclave sterilization monitoring) is NOT true?
Show answer and explanation
Correct answer: C. Spore testing should be done each day the autoclave is used
The CDC and ADA recommend that dental offices perform biological indicator (spore) testing at least weekly — not necessarily every day the sterilizer is used. Biological monitoring using Geobacillus stearothermophilus spore strips or ampoules is the gold standard for verifying sterilizer performance. A positive spore test after a sterilization cycle indicates a malfunction; the sterilizer should be taken out of service, the cause identified and corrected, and the test repeated before returning to clinical use. In-office dental sterilizers should be tested weekly.