All of the following are acceptable statements about biological spore testing EXCEPT:

Patient case: Preparing the practice

demo: New associate dentist

chief: Reviewing the practice infection control and emergency protocols before starting clinical work.

history: The practice is updating its exposure control plan and its written office emergency procedures.

meds: Not applicable.

findings: The office manager asks a series of questions about immunisation, sterilisation monitoring, waste handling and the hierarchy of hazard controls.

  1. A test should be run after any major sterilizer repair
  2. Monthly testing is sufficient for a routine sterilizer
  3. A spore test should accompany every implantable device load
  4. Testing should be performed at least weekly
Show answer and explanation

Correct answer: B. Monthly testing is sufficient for a routine sterilizer

  • A. This is true. Any repair, relocation, or misuse warrants testing, conventionally three consecutive empty-chamber tests before returning the unit to service.
  • B. Correct, and this is the exception. Monthly is not adequate. CDC guidance calls for biological monitoring at least weekly, so that a sterilizer failure is caught within days rather than weeks of instrument use.
  • C. This is true. Loads containing implantable devices should carry a biological indicator, and the device should be quarantined until the result is known.
  • D. This is true and is the accepted standard: at least weekly biological monitoring with an appropriate spore preparation.